Hardware
Weighing scale configuration
Setting an Essae DS-252 or a Mettler Toledo IND560 to stream weight over RS232 in the format the data logger reads.
The data logger reads whatever the scale sends, so the scale has to be sending weight on its own, as text, at 9600 baud. Most scales ship that way or close to it. Two that need a few settings changed are covered here. To check a scale that is not on this page, use Weighing scale compatibility.
Essae DS-252
Every setting is a numbered code, C10 to C23. You step through the codes and set each one to the value in the table.
Keys
| Key | What it does |
|---|---|
| Zero | Enters configuration mode, together with the left arrow |
| Left arrow | Moves to the tens digit |
| Up arrow | Changes the digit under the cursor, and steps to the next code |
Star (*) | Saves the value |
Enter configuration mode
- Press and hold Zero.
- While holding it, press left arrow three times in quick succession.
Change a value
- Press up arrow until the display shows the code you want, for example C10. After C25 it wraps back to C10.
- Press star to open that code.
- Up arrow changes the right-hand digit. It only counts up, and goes from 9 back to 0.
- To change the left-hand digit, press left arrow once, then use up arrow.
- Press star to save.
Settings
Set each code to its value. Write down what the scale had before you start, in case you need to put it back.
| Code | Value | Code | Value |
|---|---|---|---|
| C10 | 14 | C17 | 00 |
| C11 | 21 | C18 | 40 |
| C12 | 10 | C19 | 00 |
| C13 | 00 | C20 | 00 |
| C14 | 01 | C21 | 00 |
| C15 | 11 | C22 | 00 |
| C16 | 11 | C23 | 00 |
Mettler Toledo IND560
The IND560 has a COM1 port as standard. It is a three-wire RS232 interface: TXD, RXD and GND, with XON/XOFF flow control available. You configure what it sends with an output template, and tell COM1 to send that template continuously.
Build the template
- In the Communications menu, go to Templates > Output.
- Five templates are available, Template1 to Template5. Pick an empty one. The rest of this page uses Template3.
- Press VIEW TABLE to see its elements. Press CLEAR to remove all of them, then EXIT to leave the table.
Now fill in the table:
- Use UP/DOWN to select the template and press EDIT.
- Select the first record and set Element Type to
wt0102. - Press ENTER for more options. Set Format to
rightand Leading zero fill todisabled. Press OK. - Press DOWN to the next element and EDIT it.
- Set Element Type to
special characterand the value to0x0D(decimal 13, the carriage return\r). Press OK.
That gives one line per reading: the weight, then a carriage return.
Send it continuously
- In the connection list, use UP/DOWN to highlight COM1 and press EDIT/INSERT.
- Set Assignment to
Continuous Templateand Template toTemplate3. Press OK.
Set up COM1
On the COM1 setup screen:
| Setting | Value |
|---|---|
| Baud | 9600 |
| Data bits | 8 |
| Parity | none |
| Flow control | none |
| Interface | RS232 |
You do not need to open the panel-mount enclosure to wire COM1. It is the 8-terminal connector at the rear, at the bottom of the panel. Turn the panel over so the connector faces you. TxD, RxD and GND are the first three terminals from the left.
Test it
- Connect the terminal to a laptop with shielded twisted pair. Cross the wires, so the terminal’s TxD goes to the laptop’s RxD and the other way round.
- Open a serial terminal on the laptop with the same settings as above. The
steps for HTerm,
screenandminicomare in Weighing scale compatibility. - Power up the terminal and the scale.
- Put a weight on the platform, or press down on it and let go. The weight should scroll past on the laptop.
Reference: Mettler Toledo IND560 Terminal Technical Manual, 71209394.